移液器性能指标参数
量程 |
增量 |
测量体积 |
允许最大系统误差 |
允许最大随机误差 |
||
|
μl |
μl |
±% |
±μl |
±% |
±μl |
0.1-2.5µl |
0.05 |
0.5 |
2 |
0.01 |
2.1 |
0.0105 |
1.5 |
0.7 |
0.0105 |
1.6 |
0.024 |
||
2.5 |
0.4 |
0.01 |
1.2 |
0.03 |
||
0.5-10µl
|
0.1
|
1 |
1.0 |
0.01 |
1.8 |
0.018 |
5 |
0.4 |
0.02 |
0.4 |
0.02 |
||
10 |
0.5 |
0.05 |
1.1 |
0.11 |
||
2-20μl
|
0.5
|
2 |
0.4 |
0.008 |
0.4 |
0.008 |
10 |
0.4 |
0.04 |
0.5 |
0.05 |
||
20 |
0.5 |
0.1 |
0.4 |
0.08 |
||
5-50µl |
0.5 |
5 |
2 |
0.1 |
1.1 |
0.055 |
25 |
1.3 |
0.325 |
0.7 |
0.175 |
||
50 |
0.6 |
0.3 |
1.5 |
0.75 |
||
10-100μl
|
1
|
10 |
0.2 |
0.02 |
0.9 |
0.09 |
50 |
0.2 |
0.1 |
0.9 |
0.45 |
||
100 |
0.3 |
0.3 |
0.9 |
0.9 |
||
20-200μl
|
1
|
20 |
0.5 |
0.1 |
0.8 |
0.16 |
100 |
0.3 |
0.3 |
0.2 |
0.2 |
||
200 |
0.3 |
0.6 |
0.1 |
0.2 |
||
50-200µl |
1
|
50 |
0.2 |
0.1 |
0.8 |
0.4 |
100 |
0.3 |
0.3 |
0.2 |
0.2 |
||
200 |
0.3 |
0.6 |
0.1 |
0.2 |
||
100-1000μl
|
5
|
100 |
0.3 |
0.3 |
0.1 |
0.1 |
500 |
0.3 |
1.5 |
0.1 |
0.5 |
||
1000 |
0.3 |
3 |
0.1 |
1 |
||
200-1000µl |
5
|
200 |
0.3 |
0.6 |
0.1 |
0.2 |
500 |
0.3 |
1.5 |
0.1 |
0.5 |
||
1000 |
0.3 |
3 |
0.1 |
1 |
||
1000-5000µl |
50 |
1000 |
0.3 |
3 |
0.1 |
1 |
2500 |
0.3 |
7.5 |
0.1 |
2.5 |
||
5000 |
0.3 |
15 |
0.1 |
5 |
||
2-10ml |
100 |
2000 |
0.4 |
8 |
0.1 |
2 |
5000 |
0.4 |
20 |
0.1 |
5 |
||
10000 |
0.4 |
40 |
0.1 |
10 |
||
0.5-10µl(八通道)
|
0.1
|
1 |
1.0 |
0.01 |
1.8 |
0.018 |
5 |
0.4 |
0.02 |
0.4 |
0.02 |
||
10 |
0.5 |
0.05 |
1.1 |
0.11 |
||
5-50µl(八通道)
|
0.5 |
5 |
2 |
0.1 |
1.1 |
0.055 |
25 |
1.3 |
0.325 |
0.7 |
0.175 |
||
50 |
0.6 |
0.3 |
1.5 |
0.75 |
||
50-300µl (八通道)
|
5
|
50 |
0.2 |
0.1 |
0.8 |
0.4 |
150 |
0.3 |
0.45 |
0.2 |
0.3 |
||
300 |
0.3 |
0.9 |
0.1 |
0.3 |
标签: 半只消毒,量程:20-200μl